A bench-top megavoltage fan-beam CT using CdWO4-photodiode detectors. II. Image performance evaluation
- 1. Department of Medical Physics, Cross Cancer Institute and Department of Oncology, University of Alberta, 11560 University Avenue, Edmonton, Alberta T6G 1Z2 (Canada)
- 2. Department of Medical Physics, Cross Cancer Institute and Departments of Oncology and Physics, University of Alberta, 11560 University Avenue, Edmonton, Alberta T6G 1Z2 (Canada)
Description
Megavoltage computed tomography (MVCT) is a potential imaging tool for positioning and dose delivery verification during image guided radiotherapy. The problem with many MVCT detectors, however, is their low detective quantum efficiency (DQE) which leads to poor low contrast resolution (LCR) and high image noise. This makes separating the tumors from the soft tissue background difficult. This manuscript describes the imaging performance of our bench-top MVCT scanner that uses an 80-element detector array consisting of CdWO4-photodiode elements with a DQE of 19% in 6 MV and 26% in Co60 beams [T. T. Monajemi, S. Steciw, B. G. Fallone, and S. Rathee, 'Modelling scintillator-photodiodes as detectors for megavoltage CT', Med. Phys. 31, 1225-1234 (2004)] at zero frequency. The imaging experiments presented were carried out mainly in a Co60 teletherapy unit, while the beam hardening characteristics of the system were also presented for a 6 MV beam. During image evaluation, persistent ring artifacts, caused by air gaps at the ends of the eight-element detector blocks, were removed by using a calibration procedure. The measured contrast of a low contrast target with a 20 mm diameter was determined to be independent of dose, between 2.1 and 17 cGy. The measured LCR of a target with a nominal contrast of 2.8% was reduced from 2.3% to 1.2% when the contrast target diameter was reduced from 15 to 5 mm, using 17 cGy for imaging. The signal to noise ratio of this system is shown to be proportional to the square root of dose. Most importantly, a low contrast target with a diameter of 6 mm and a nominal contrast level of 1.5% is resolved with a radiation dose of 2.1 cGy in the Co60 beam. The spatial resolution in the Co60 beam is limited to one line pair per centimeter mainly due to the size of the Co60 source
Additional details
Identifiers
- DOI
- 10.1118/1.2181291;
Publishing Information
- Journal Title
- Medical Physics
- Journal Volume
- 33
- Journal Issue
- 4
- Journal Page Range
- p. 1090-1100
- ISSN
- 0094-2405
- CODEN
- MPHYA6
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 37078680
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- CADMIUM TUNGSTATES; COBALT 60; COMPUTERIZED TOMOGRAPHY; EVALUATION; IMAGE SCANNERS; IMAGES; PERFORMANCE; PHOTODIODES; QUANTUM EFFICIENCY; RADIATION DOSES; RADIOTHERAPY; SIGNAL-TO-NOISE RATIO; SPATIAL RESOLUTION
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; CADMIUM COMPOUNDS; COBALT ISOTOPES; DIAGNOSTIC TECHNIQUES; DIMENSIONLESS NUMBERS; DOSES; EFFICIENCY; INORGANIC PHOSPHORS; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTOPES; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; MEDICINE; MINUTES LIVING RADIOISOTOPES; NUCLEAR MEDICINE; NUCLEI; ODD-ODD NUCLEI; OXYGEN COMPOUNDS; PHOSPHORS; RADIOISOTOPES; RADIOLOGY; REFRACTORY METAL COMPOUNDS; RESOLUTION; SEMICONDUCTOR DEVICES; SEMICONDUCTOR DIODES; THERAPY; TOMOGRAPHY; TRANSITION ELEMENT COMPOUNDS; TUNGSTATES; TUNGSTEN COMPOUNDS; YEARS LIVING RADIOISOTOPES
Optional Information
- Notes
- (c) 2006 American Association of Physicists in Medicine