Description

Six base of skull IMRT treatment plans were delivered to 3D dosimeters within the RPC Head and Neck Phantom for QA verification. Isotropic 2mm 3D data was obtained using the DLOS-PRESAGE system and compared to an Eclipse (Varian) treatment plan. Normalized Dose Distribution pass rates were obtained for a number of criteria. High quality 3D dosimetry data was observed from the DLOS system, illustrated here through colormaps, isodose lines, profiles, and NDD 3D maps. Excellent agreement with the planned dose distributions was also observed with NDD analysis revealing > 90% NDD pass rates [3%, 2mm], noise < 0.5%. This paper focuses on a detailed exploration of the quality and use of 3D dosimetry data obtained with the DLOS-PRESAGE system.

Availability note (English)

Available from http://dx.doi.org/10.1088/1742-6596/444/1/012049

Additional details

Publishing Information

Journal Title
Journal of Physics. Conference Series (Online)
Journal Volume
444
Journal Issue
1
Journal Page Range
[4 p.]
ISSN
1742-6596

Conference

Title
7. international conference on 3D radiation dosimetry
Acronym
IC3DDose
Dates
4-8 Dec 2012
Place
Sydney (Australia)

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
44074769
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
DOSEMETERS; DOSIMETRY; HEAD; MAPS; NECK; PHANTOMS; RADIATION DOSE DISTRIBUTIONS; SKULL; VERIFICATION
Descriptors DEC
BODY; MEASURING INSTRUMENTS; MOCKUP; ORGANS; SKELETON; STRUCTURAL MODELS