Demodulator electronics for laser vibrometry
Creators
- 1. Laser and Fibre Electronics Group, Wroclaw University of Technology, Wybrzeze Wyspianskiego 27, 50-370 Wroclaw (Poland)
Description
One of the most important parts of a fiber-laser vibrometer is demodulation electronic section. The distortion, nonlinearity, offset and added noise of measured signal come from electronic circuits and they have direct influence on finale measuring results. Two main parameters of an investigated vibrating object: velocity V(t) and displacement s(t), influence of detected beat signals. They are: the Doppler frequency deviation f(t) and phase shift φ(t), respectively. Because of wide range of deviations it is difficult to use just one demodulator. That is the reason why we use three different types of demodulators. The first one is the IQ demodulator, which is the most sensitive one and its output is proportional to the displacement. Each IQ channel is sampled simultaneously by an analog to digital converter (ADC) integrated in a digital signal processor (DSP). The output signals from the two FM demodulators are proportional to the frequency deviation of heterodyne signals. They are sensitive directly to the velocity of the object. The main disadvantage of scattered light interferometry system is a ''speckle effect'', appearing in relatively large amplitude fluctuation of a heterodyne signal. To minimize ''speckle effect'' influence on quality of beat signals we applied the automatic gain control (AGC) system. Data acquisition, further signal processing (e.g. vibration frequency spectra) and presentation of results is realized by PC via USB interface.
Additional details
Identifiers
- DOI
- 10.1063/1.4730540;
Publishing Information
- Journal Title
- AIP Conference Proceedings
- Journal Volume
- 1457
- Journal Issue
- 1
- Journal Page Range
- p. 35-40
- ISSN
- 0094-243X
- CODEN
- APCPCS
Conference
- Title
- 10. international conference on vibration measurements by laser and noncontact techniques
- Acronym
- AIVELA 2012
- Dates
- 27-29 Jun 2012
- Place
- Ancona (Italy)
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 43094287
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ANALOG-TO-DIGITAL CONVERTERS; DATA ACQUISITION; DATA PROCESSING; DOPPLER EFFECT; ELECTRONIC CIRCUITS; EQUIPMENT INTERFACES; GAIN; INTERFEROMETRY; LASERS; MICROCOMPUTERS; NONLINEAR PROBLEMS; OPTICAL FIBERS; PHASE SHIFT; SIGNALS
- Descriptors DEC
- AMPLIFICATION; COMPUTERS; DIGITAL COMPUTERS; ELECTRONIC EQUIPMENT; EQUIPMENT; FIBERS; PROCESSING
Optional Information
- Notes
- (c) 2012 American Institute of Physics