Study on the relationship between water-to-detection crystal distance and detection efficiency in underwater γ spectrum measurement
- 1. Naval Characteristic Medical Center, Shanghai (China)
Description
The quantitative relationship between the water to detection crystal distance and the detector efficiency must be determined in the design of the shell dimension of the underwater gamma spectrometer. In this study, a 3-inch LaBr3 crystal and hollow water sphere model was established based on the underwater γ spectrum measurement conditions. The γ-ray counting rate that reached the surface of the detector crystal and γ ray full-energy peak counting rate were calculated respectively through mathematical calculation and Monte Carlo sampling calculation. The results show that the detection efficiency of the specific γ ray full-energy peak emitted from the water is constant while the distance between the detection crystal and the water is changing. The conclusion provides a basis for the design of the shell structure of the underwater gamma spectrometer. (authors)
Additional details
Publishing Information
- Journal Title
- Nuclear Electronics and Detection Technology
- Journal Volume
- 40
- Journal Issue
- 2
- Journal Page Range
- p. 330-334
- ISSN
- 0258-0934
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 55086782
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- COUNTING RATES; CRYSTALS; DETECTION; DISTANCE; EFFICIENCY; GAMMA SPECTROMETERS; LANTHANUM BROMIDES; MONTE CARLO METHOD; PEAKS; SAMPLING; SPECTRA; SURFACES; UNDERWATER; WATER
- Descriptors DEC
- BROMIDES; BROMINE COMPOUNDS; CALCULATION METHODS; HALIDES; HALOGEN COMPOUNDS; HYDROGEN COMPOUNDS; LANTHANUM COMPOUNDS; LANTHANUM HALIDES; LEVELS; MEASURING INSTRUMENTS; OXYGEN COMPOUNDS; RARE EARTH COMPOUNDS; SPECTROMETERS
Optional Information
- Notes
- 4 figs., 1 tab., 7 refs.