Published February 1988 | Version v1
Journal article

New position sensitive photomultiplier tubes for high energy physics and nuclear medical applications

  • 1. Hamamatsu Photonics K.K., 314-5, Shimokanzo, Toyooka Village, Iwata-Gun, Sizuoka Pref. 438-01 (Japan)

Description

New position sensitive photomultiplier tubes with fine mesh structured dynodes and discrete anode array configurations have been developed. One kind of the position sensitive photomultiplier tubes is being used as a photodetector for High Enegy Physics applications in high magnetic field environments. A photomultiplier tube constructed with 88 Multi-Anodes has a spatial resolution of less than 2.6 mm in FWHM in a magnetic field with a density of 500-2000 Gauss. The resolution includes an anode width of 2.6 mm. Another type of Multi-Anode photomultiplier tube which has been developed is the detector with a high spatial resolution for such applications as the PET system and hodoscope in scintillation systems. The tube, by applying additional electro-focusing electrodes, has an intrinsic spatial resolution of 1.4 mm in FWHM without the magnetic field

Additional details

Publishing Information

Journal Title
IEEE Trans. Nucl. Sci.
Journal Volume
35
Journal Issue
1
Series
IEEE Trans. Nucl. Sci.
Journal Page Range
382-386
ISSN
0018-9499
CODEN
IETNA

Conference

Title
34. nuclear science symposium and 19. nuclear power systems symposium.
Dates
21-23 Oct 1987.
Place
San Francisco, CA (USA).

Optional Information

Secondary number(s)
CONF-871006--.