Creators
- 1. Department of Radiation Oncology, Homi Bhabha Cancer Hospital and Research Centre, Visakhapatnam, Andhra Pradesh (India)
- 2. Department of Radiation Oncology, Tata Medical Centre, Kolkata, West Bengal (India)
Description
To commission and validate commercial deformable image registration (DIR) systems (SmartAdapt® and Velocity™ using task group 132 (TG-132) digital phantom datasets. Additionally, the study compares and verifies the DIR algorithms of the two systems.The DIR algorithms of SmartAdapt and Velocity were commissioned and their deformation results were compared. Both systems can be used for clinical purpose. While there were only minimal differences between the two systems, Velocity provided lower values for parotids, bladder, rectum, and prostate (soft tissue) compared to SmartAdapt. However, for mandible, spinal cord, and femoral heads (rigid structures), both systems showed nearly identical results
Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of Medical Physics
- Journal Volume
- 49
- Journal Issue
- 2
- Journal Page Range
- p. 240-249
- CODEN
- JMPHFE
INIS
- Country of Publication
- India
- Country of Input or Organization
- India
- INIS RN
- 55072738
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- CAT SCANNING; IMAGE PROCESSING; NEOPLASMS; PERFORMANCE; PERFORMANCE TESTING; PROSTATE; RADIOTHERAPY
- Descriptors DEC
- BODY; COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; DISEASES; GLANDS; MALE GENITALS; MEDICINE; NUCLEAR MEDICINE; ORGANS; PROCESSING; RADIOLOGY; TESTING; THERAPY; TOMOGRAPHY