Published December 2003
| Version v1
Journal article
Dose prediction and process optimization in a gamma sterilization facility using 3-D Monte Carlo code
Creators
Description
A model of a gamma sterilizer was built using the ITS/ACCEPT Monte Carlo code and verified through dosimetry. Individual dosimetry measurements in homogeneous material were pooled to represent larger bodies that could be simulated in a reasonable time. With the assumptions and simplifications described, dose predictions were within 2-5% of dosimetry. The model was used to simulate product movement through the sterilizer and to predict information useful for process optimization and facility design
Additional details
Additional titles
- Augmented title (English)
- Monte Carlo; Gamma; Simulation; Dosimetry; Verification; ITS; Mathematical model; Modeling
Identifiers
- DOI
- 10.1016/j.radphyschem.2003.09.001;
- PII
- S0969806X03005164;
Publishing Information
- Journal Title
- Radiation Physics and Chemistry (1993)
- Journal Volume
- 68
- Journal Issue
- 6
- Journal Page Range
- p. 947-958
- ISSN
- 0969-806X
- CODEN
- RPCHDM
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- Bulgaria
- INIS RN
- 35033422
- Subject category
- S38: RADIATION CHEMISTRY, RADIOCHEMISTRY AND NUCLEAR CHEMISTRY;
- Descriptors DEI
- GAMMA DOSIMETRY; GAMMA RADIATION; MONTE CARLO METHOD; OPTIMIZATION; RADIATION DOSES; RADIOSTERILIZATION; VERIFICATION
- Descriptors DEC
- CALCULATION METHODS; DOSES; DOSIMETRY; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; IRRADIATION; RADIATIONS; STERILIZATION
Optional Information
- Copyright
- Copyright (c) 2003 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.