Published July 7, 2005 | Version v1
Report

Combining a thermal-imaging diagnostic with an existing imaging VISAR diagnostic at the National Ignition Facility (NIF)

Description

Optical diagnostics are currently being designed to analyze high-energy density physics experiments at the National Ignition Facility (NIF). Two independent line-imaging Velocity Interferometer System for Any Reflector (VISAR) interferometers have been fielded to measure shock velocities, breakout times, and emission of targets having sizes of 1-5 mm. An 8-inch-diameter, fused silica triplet lens collects light at f/3 inside the 30-foot-diameter NIF vacuum chamber. VISAR recordings use a 659.5-nm probe laser. By adding a specially coated beam splitter to the interferometer table, light at wavelengths from 540 to 645 nm is spilt into a thermal-imaging diagnostic. Because fused silica lenses are used in the first triplet relay, the intermediate image planes for different wavelengths separate by considerable distances. A corrector lens on the interferometer table reunites these separated wavelength planes to provide a good image. Thermal imaging collects light at f/5 from a 2-mm object placed at Target Chamber Center (TCC). Streak cameras perform VISAR and thermal-imaging recording. All optical lenses are on kinematic mounts so that pointing accuracy of the optical axis may be checked. Counter-propagating laser beams (orange and red) are used to align both diagnostics. The red alignment laser is selected to be at the 50 percent reflection point of the beam splitter. This alignment laser is introduced at the recording streak cameras for both diagnostics and passes through this special beam splitter on its way into the NIF vacuum chamber

Availability note (English)

Available from PURL: https://www.osti.gov/servlets/purl/15016768-HpMyNy/native/

Additional details

Publishing Information

Imprint Pagination
[vp.]
Report number
UCRL-CONF--213575

Conference

Title
2005 SPIE Optics and Photonics Conference
Dates
31 Jul - 4 Aug 2005
Place
San Diego, CA (United States)

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
36101191
Subject category
S70: PLASMA PHYSICS AND FUSION TECHNOLOGY;
Resource subtype / Literary indicator
Conference, Non-conventional Literature
Descriptors DEI
ACCURACY; ALIGNMENT; DIAGNOSTIC TECHNIQUES; INTERFEROMETERS; LASERS; LENSES; PLASMA DIAGNOSTICS; SILICA; STREAK CAMERAS; TARGET CHAMBERS; US NATIONAL IGNITION FACILITY
Descriptors DEC
ACCELERATOR FACILITIES; CAMERAS; MEASURING INSTRUMENTS; MINERALS; OXIDE MINERALS

Optional Information

Contract/Grant/Project number
W--7405-ENG-48
Notes
0.7 Megabytes
Funding organization
US Department of Energy (United States)