Description
Currently the Gamma Knife system is accompanied with a treatment planning system, Leksell GammaPlan (LGP) which is a standard, computer-based treatment planning system for Gamma Knife radiosurgery. In LGP, the dose calculation algorithm does not consider the scatter dose contributions and the inhomogeneity effect due to the skull and air cavities. To improve the dose calculation accuracy, Monte Carlo simulations have been implemented for the Gamma Knife planning system. In this work, the 201 Cobalt-60 sources in the Gamma Knife unit are considered to have the same activity. Each Cobalt-60 source is contained in a cylindric stainless steel capsule. The particle phase space information is stored in four beam data files, which are collected in the inner sides of the 4 treatment helmets, after the Cobalt beam passes through the stationary and helmet collimators. Patient geometries are rebuilt from patient CT data. Twenty two Patients are included in the Monte Carlo simulation for this study. The dose is calculated using Monte Carlo in both homogenous and inhomogeneous geometries with identical beam parameters. To investigate the attenuation effect of the skull bone the dose in a 16cm diameter spherical QA phantom is measured with and without a 1.5mm Lead-covering and also simulated using Monte Carlo. The dose ratios with and without the 1.5mm Lead-covering are 89.8% based on measurements and 89.2% according to Monte Carlo for a 18mm-collimator Helmet. For patient geometries, the Monte Carlo results show that although the relative isodose lines remain almost the same with and without inhomogeneity corrections, the difference in the absolute dose is clinically significant. The average inhomogeneity correction is (3.9 ± 0.90) % for the 22 patients investigated. These results suggest that the inhomogeneity effect should be considered in the dose calculation for Gamma Knife treatment planning
Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of Physics. Conference Series (Online)
- Journal Volume
- 74
- Journal Issue
- 1
- Journal Page Range
- p. 021023
- ISSN
- 1742-6596
Conference
- Title
- 1. European workshop on Monte Carlo treatment planning
- Dates
- 22-25 Oct 2006
- Place
- Gent (Belgium)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 38080221
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ACCURACY; ALGORITHMS; COBALT 60; COMPUTERIZED SIMULATION; CORRECTIONS; DOSES; GAMMA RADIATION; MONTE CARLO METHOD; PATIENTS; PHANTOMS; PHASE SPACE; PLANNING; QUALITY ASSURANCE; RADIOTHERAPY; SKULL; STAINLESS STEELS
- Descriptors DEC
- ALLOYS; BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; BODY; CALCULATION METHODS; CARBON ADDITIONS; COBALT ISOTOPES; ELECTROMAGNETIC RADIATION; HIGH ALLOY STEELS; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTOPES; IONIZING RADIATIONS; IRON ALLOYS; IRON BASE ALLOYS; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; MATHEMATICAL LOGIC; MATHEMATICAL SPACE; MEDICINE; MINUTES LIVING RADIOISOTOPES; MOCKUP; NUCLEAR MEDICINE; NUCLEI; ODD-ODD NUCLEI; ORGANS; RADIATIONS; RADIOISOTOPES; RADIOLOGY; SIMULATION; SKELETON; SPACE; STEELS; STRUCTURAL MODELS; THERAPY; TRANSITION ELEMENT ALLOYS; YEARS LIVING RADIOISOTOPES