Investigation of the leakage radiation dose rate from an auto infusion device in 18F FDG PET
- 1. Ogaki Municipal Hospital, Department of Medical Technology, Ogaki, Gifu (Japan)
Description
The aim of this study was to investigate the leakage radiation dose rate on the surface of an auto infusion device in 18F-fludeoxyglucose (FDG) positron emission tomography. We preliminary measured the radioactivity of delivery-supplied FDG drugs. Leakage radiation doses on the six (front, rear, right, left, top, and bottom) surface points of the auto infusion device were measured ninety minutes before the assay time using an energy-compensated gamma scintillation survey meter and an ionization chamber-type survey meter. Leakage radiation doses on the front, rear, right, left, and top surfaces were less than 7μSv/h and satisfied specification of the auto infusion device. However, a leakage dose exceeding 40 μSv/h was detected on the bottom surface. (author)
Abstract (Japanese)
デリバリー供給18F-fludeoxyglucose(FDG)によるpositron emission tomography検査において,放射性薬剤自動投与装置の筐体表面6カ所の漏洩線量率を調査した.デリバリー供給FDG製剤の放射能量を実測し,検定時刻の90分前に放射性薬剤自動投与装置にFDG製剤をセットした状態で筐体の前後,左右および上下面の漏洩線量率をγ線用シンチレーションサーベイメータおよび電離箱式サーベイメータにて計測した.デリバリー供給FDG製剤の放射能量は326.8-333.7 MBqであった.筐体の前後左右および上面における漏洩線量率は7 µSv/h以下であり装置仕様を満たしていた.筐体下面では40 µSv/hを超過する漏洩線量率が確認された.(著者)Additional details
Additional titles
- Original title (Japanese)
- 18F-FDG PET検査における放射性薬剤自動投与装置筐体表面の漏洩線量率についての調査
Identifiers
Publishing Information
- Journal Title
- Nippon Hoshasen Gijutsu Gakkai Zasshi (Online)
- Journal Volume
- 79
- Journal Issue
- 8
- Series
- 雑誌名:日本放射線技術学会雑誌
- Journal Page Range
- p. 818-823
- ISSN
- 1881-4883
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 55065255
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- DOSE RATES; FLUORINE 18; FLUORODEOXYGLUCOSE; INFUSION; LEAK TESTING; LEAKS; OCCUPATIONAL EXPOSURE; POSITRON COMPUTED TOMOGRAPHY; RADIOACTIVITY; SURFACE CONTAMINATION
- Descriptors DEC
- ANTIMETABOLITES; BETA DECAY RADIOISOTOPES; BETA-PLUS DECAY RADIOISOTOPES; COMPUTERIZED TOMOGRAPHY; CONTAMINATION; DIAGNOSTIC TECHNIQUES; DRUGS; EMISSION COMPUTED TOMOGRAPHY; FLUORINE ISOTOPES; HOURS LIVING RADIOISOTOPES; INTAKE; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; LIGHT NUCLEI; NANOSECONDS LIVING RADIOISOTOPES; NUCLEI; ODD-ODD NUCLEI; RADIOISOTOPES; TESTING; TOMOGRAPHY
Optional Information
- Notes
- 21 refs., 2 figs., 1 tab.