Published November 2007 | Version v1
Journal article

Tow efficiency correction functions of source self-absorption of an HPGe detector

  • 1. Submarine Academy, Qingdao (China)

Description

The efficiency correction function of source absorption of an HPGe γ detector is determined by experiment in energy range from 59.5 keV to 1408 keV and density range from 0.3 g/cm3 to 2.0 g/cm3. Fit Polynomial and fit Sigmoidal are compared. The results show that fit Sigmoidal is better than fit polynomial, and the detection efficiency at any points of energy and density could be conveniently calculated with it in calibrated range. (authors)

Additional details

Publishing Information

Journal Title
Nuclear Electronics and Detection Technology
Journal Volume
27
Journal Issue
6
Journal Page Range
p. 1083-1086
ISSN
0258-0934

Optional Information

Notes
6 figs., 2 tabs., 2 refs.