Serial measurement of radiation leakage dose rates in safekeeping at the Gammaknife room

  • 1. Kanto Medical Center NTT EC, Tokyo (Japan)

Description

We report the serial measurement of leakage dose rates in safekeeping at the Gammaknife room during the past 4 years and 9 months by scintillation survey meter. The leakage dose rates at the radiation boundaries were the same as the natural background levels. Leakage dose rates at each shield calculation point from two 90Sr calibration sources contained in the storehouse were negligible compared with those from 60Co sources of the Gammaknife. 60Co sources of the Gammaknife are arranged in 201 pieces at 10 degree interval on the circumference and in five lines within an arc of 35 degrees. Its shield container is made of iron at least 43 cm thick. We got leakage dose rates less than 40% of the calculated values. We think it is caused by the difference of each actual distance and shield thickness because 60Co sources are usually considered as a point source in the shield calculation. There are shutters opening up and down when patients go in and out to the direction of the couch. The leakage values to this direction were about twice as much as the calculated value. So, we knew the thickness of those shutters was thinner than 43 cm. The half life time of 60Co source calculated from the serial measurements of leakage dose rates was 4.93 years on average. It is 94% of the physical half life value of 5.27 years. We judged it was acceptable considering the difficulty of measuring low dose rate level with the radiation survey meter. Very strong correlation was observed between the decrease of 60Co dose rate acquired from one minute measurement at the center of 18 cm diameter polysterene phantom gotten from December 2000 to August 2005 and that of computation based on the physical half life time. Likewise there was strong and more correlation with leakage dose rate in the Gammaknife room. From this, we deduce the leakage dose rate decreases according to the theory of the disintegration of radioactivity with passage of time. Revised radiation related laws took effect from June 2005 and regulated the receipt and expenditure of radioisotopes definitely, so we have to record those articles strictly. From this cause, we can grasp the radioactive disintegration more surely than before. It is unreasonable we can not deal with the disintegration in caluculating usage time of capsuled radioisotopes. We hope we can handle the correction of the disintegraton of capsuled radioisotopes as well as uncapsuled ones based on the result of this investigation about serial measurement of leakage dose rate in safekeeping. (author)

Additional details

Publishing Information

Journal Title
Nippon Hoshasen Gishikai Zasshi
Journal Volume
53
Journal Issue
9
Journal Page Range
p. 943-951
ISSN
0287-9395