Published 2003 | Version v1
Report Open

Ion feedback afterpulses in FEU-130 and XP2020 photomultipliers

Description

Intensities and time distributions of ion feedback afterpulses are studied as a function of the difference of potentials between the photocathode and the first dynodes in FEU-130. The intensities and amplitudes of afterpulses in FEU-130 are found to be appreciably larger than in XP2020, which may limit the use of FEU-130 in some experiments with autocorrelation delayed-coincidence spectrometers

Availability note (English)

Available from INIS in electronic form

Files

35055727.pdf

Files (466.7 kB)

Name Size Download all
md5:721e5f0af1be1e5db898edd276249bf0
466.7 kB Preview Download

System files (27.0 kB)

Name Size Download all

Additional details

Additional titles

Original title (Russian)
Послеимпульсы ионной обратной связи в фотоумножителях ФЭУ-130 и XP2020

Publishing Information

Imprint Pagination
24 p.
Report number
JINR-R--13-2003-242

INIS

Country of Publication
Joint Institute for Nuclear Research (JINR)
Country of Input or Organization
Joint Institute for Nuclear Research (JINR)
INIS RN
35055727
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
COINCIDENCE SPECTROMETRY; ELECTRIC POTENTIAL; EXCITED STATES; FEEDBACK; LIFETIME; PHOTOMULTIPLIERS
Descriptors DEC
COINCIDENCE METHODS; COUNTING TECHNIQUES; ENERGY LEVELS; PHOTOTUBES

Optional Information

Notes
6 refs., 15 figs. Submitted to the journal, Pribory i Tekhnika Ehksperimenta