Published 1997 | Version v1
Journal article

Development of a rapid procedure for the measurement of uranium in drinking water by PERALSR spectrometry

  • 1. ETRAC Laboratories, Oak Ridge TN (United States)
  • 2. Radiation Safety Engineering, Chandler AZ (United States)
  • 3. Tennessee Univ., Knoxville, TN (United States). Dept. of Chemistry

Description

A new method has been developed for the measurement of U in water samples by selective and quantitative extraction from DTPA solution into an extractive scintillator containing HDEHP and alpha-counting by PERALSR spectrometry. Because this method offers several advantages over the current EPA and ASTM standard test methods for U in drinking water, including speed, simplicity and isotopic information, it has been proposed as a new ASTM standard test method. Sample preparation requires from 1-4 h. Less than 0.5% of Am, Pu, Po, Ra and Th were found to extract under the prescribed conditions. Typical backgrounds are ≤ 0.01 cpm. A comparison of this method with EPA standard method 908.0 and the results of two interlaboratory test of this method are reported. (author)

Additional details

Publishing Information

Journal Title
Journal of Radioanalytical and Nuclear Chemistry
Journal Volume
221
Journal Issue
1-2
Journal Page Range
p. 115-122.
ISSN
0236-5731
CODEN
JRNCDM

Optional Information

Notes
12 refs.