Published 1985
| Version v1
Report
Open
8 channel gas threshold Cherenkov counter
- 1. Joint Inst. for Nuclear Research, Dubna (USSR). Lab. of Nuclear Problems
- 2. Gosudarstvennyj Komitet po Ispol'zovaniyu Atomnoj Ehnergii SSSR, Serpukhov. Inst. Fiziki Vysokikh Ehnergij
Description
A wide aperture 8-channel gas threshold Cherenkov counter used in HYPERON spectrometric complex which is designed for inR-beam investigations at the Serpukhov synchrotron is described. The radiator dimensions are 240x240x95 cm, the operating pressure is 1 atm. Eight thin rectangular spherical mirrors 76x31 cm and photomultipliers of XP 2041/Q are used. Testing results are given The velocity resolution is 3x10-5
Availability note (English)
MF available from INIS under the Report Number.Files
16078340.pdf
Files
(202.6 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:8eb08b6f29069d6aa7523180fab89fd8
|
202.6 kB | Preview Download |
System files
(26.7 kB)
| Name | Size | Download all |
|---|
Additional details
Additional titles
- Original title (Russian)
- Восьмиканальный газовый пороговый черенковский счетчик
Publishing Information
- Imprint Pagination
- 6 p.
- Report number
- JINR--13-85-80
INIS
- Country of Publication
- USSR
- Country of Input or Organization
- USSR
- INIS RN
- 16078340
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CHERENKOV COUNTERS; EFFICIENCY; ENERGY RESOLUTION; FABRICATION; GEV RANGE 01-10; KAON REACTIONS; PERFORMANCE; PHOTOMULTIPLIERS; SPECTROMETERS
- Descriptors DEC
- ENERGY RANGE; GEV RANGE; HADRON REACTIONS; MEASURING INSTRUMENTS; MESON REACTIONS; NUCLEAR REACTIONS; PHOTOTUBES; RADIATION DETECTORS; RESOLUTION
Optional Information
- Notes
- 15 refs.; 3 figs.; submitted to the journal Instrum. Tech. Exp.