Description

Monte Carlo (MC) detector simulations and measurements have been used to investigate the corrections needed for small field output factor measurements with the IRIS collimator on a CyberKnife System for a range of detectors. These corrections are found to be larger than 2.5% for all detectors at the smallest (5 mm) field size and to follow similar trends as those observed previously for fixed collimators on this treatment device. The correction factors for microchambers (Exradin A16 and PTW PinPoint 31014) ranged from 1.102 at 5 mm to 1.006 at 12.5 mm field sizes. The correction factors for diode detectors (PTW 60012 and Sun Nuclear Edge Detector) varied between 0.958 and 0.991 over the same field size range, and for a liquid filled ionization chamber (PTW microLion), the correction factor variation was 1.027-0.993. Before correction, measured output factors differed by up to 15% at 5 mm field size, which reduced to a maximum inter-detector variation of 3% after corrections were applied. The agreement of measured output factors to those calculated directly by MC simulation of dose to water was also improved after corrections were applied. In conclusion, these results support the use of measurement correction factors for small field output factor commissioning and QA. The correction factor dataset presented here can be used to apply this method to the IRIS Collimator on the CyberKnife System. (author)

Part of:
Standards, Applications and Quality Assurance in Medical Radiation Dosimetry (IDOS). Proceedings of an International Symposium. V. 1

Additional details

Publishing Information

Publisher
IAEA
Imprint Place
Vienna (Austria)
ISBN
978-92-0-116210-6
Imprint Title
Standards, Applications and Quality Assurance in Medical Radiation Dosimetry (IDOS). Proceedings of an International Symposium. V. 1
Imprint Pagination
467 p.
Series
Proceedings Series
Journal Page Range
p. 181-192
ISSN
0074-1884

Conference

Title
International Symposium on Standards, Applications and Quality Assurance in Medical Radiation Dosimetry
Acronym
IDOS 2010
Dates
9-12 Nov 2010
Place
Vienna (Austria)

INIS

Country of Publication
Austria
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
43009117
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
BEAMS; COLLIMATORS; IONIZATION CHAMBERS; MONTE CARLO METHOD; RADIOTHERAPY; SIMULATION
Descriptors DEC
CALCULATION METHODS; MEASURING INSTRUMENTS; MEDICINE; NUCLEAR MEDICINE; RADIATION DETECTORS; RADIOLOGY; THERAPY

Optional Information

Notes
4 refs., 3 tabs., 7 refs. Imprint:In two volumes; A CD-ROM of the contributed papers accompanies these proceedings.
Secondary number(s)
STI/PUB--1514(vol.1); IAEA-CN--182