Published 1997 | Version v1
Journal article

Determination of lead associated with airborne particulate matter by flame atomic absorption and wave-length dispersive x-ray fluorescence spectrometry

Creators

  • 1. Department of Chemistry, University of Isfahan, Isfahan (Iran, Islamic Republic of)

Description

The lead content of airborne particulate matter was determined by flame atomic absorption spectrometry (FAAS) following digestion with a mixture of nitric acid and hydrogen peroxide and also by wave-length dispersive x-ray fluorescence (WDXRF). The extraction procedure was checked by analyzing a standard reference material of airborne particulate matter (NIST, SRM -1648). It was concluded that lead can quantitatively (98%) be extracted from airborne particulate matter by the leaching process. A five-stage sequential extraction was performed to assess the potential mobility of lead associated with airborne particulate matter. Comparison of the airborne particulate lead measured by WDXRF to that measured by FAAS showed good agreement. The WDXRF method requires no time-consuming sample preparation or use of environmentally unfriendly solvents. The technique is suggested for direct determination of lead in airborne particulate matter in air pollution studies. (author)

Additional details

Publishing Information

Journal Title
Pollution Research
Journal Volume
16
Journal Issue
4
Journal Page Range
p. 203-207.
CODEN
PORSDX

Optional Information

Notes
15 refs., 3 figs., 4 tabs.