Published 2011 | Version v1
Journal article

A method of identification of abnormal glow curves in individual monitoring using CaSo4: Dy Teflon TLD and hot gas reader

  • 1. Radiological Physics and Advisory Div., Bhabha Atomic Research Centre (BARC), Mumbai 400 085 (India)

Description

In the present study, a method of identifying abnormal glow curves to electronically screen the glow curves of TL readout is presented. The method is based on the fact that the shape of an abnormal glow curve differs from the shape of a normal one. A few criteria for defining the normal shape of glow curves are arrived at by analysing the glow curves of dosemeters exposed to various doses in laboratory conditions and read at different elapsed time post irradiation. About 300 glow curves of dosemeters used for monthly monitoring were analysed as per these criteria and the effectiveness of the method is observed for total counts as low as 150 μSv equivalent. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/ncq549

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
144
Journal Issue
1-4
Journal Page Range
p. 195-198
ISSN
0144-8420

Conference

Title
European conference on individual monitoring of ionising radiation
Acronym
IM2010
Dates
8-12 Mar 2010
Place
Athens (Greece)

Optional Information

Notes
6 refs