Published April 21, 2009 | Version v1
Journal article

The risk of developing a second cancer after receiving craniospinal proton irradiation

  • 1. Department of Radiation Physics, University of Texas M D Anderson Cancer Center, 1515 Holcombe Blvd, Houston, TX, 77030 (United States)
  • 2. Department of Radiation Oncology, University of Texas M D Anderson Cancer Center, 1515 Holcombe Blvd, Houston, TX, 77030 (United States)
  • 3. Department of Radiation Oncology, Washington University School of Medicine, 4921, Parkview Place, St. Louis, MO 63110 (United States)

Description

The purpose of this work was to compare the risk of developing a second cancer after craniospinal irradiation using photon versus proton radiotherapy by means of simulation studies designed to account for the effects of neutron exposures. Craniospinal irradiation of a male phantom was calculated for passively-scattered and scanned-beam proton treatment units. Organ doses were estimated from treatment plans; for the proton treatments, the amount of stray radiation was calculated separately using the Monte Carlo method. The organ doses were converted to risk of cancer incidence using a standard formalism developed for radiation protection purposes. The total lifetime risk of second cancer due exclusively to stray radiation was 1.5% for the passively scattered treatment versus 0.8% for the scanned proton beam treatment. Taking into account the therapeutic and stray radiation fields, the risk of second cancer from intensity-modulated radiation therapy and conventional radiotherapy photon treatments were 7 and 12 times higher than the risk associated with scanned-beam proton therapy, respectively, and 6 and 11 times higher than with passively scattered proton therapy, respectively. Simulations revealed that both passively scattered and scanned-beam proton therapies confer significantly lower risks of second cancers than 6 MV conventional and intensity-modulated photon therapies.

Availability note (English)

Available from http://dx.doi.org/10.1088/0031-9155/54/8/002

Additional details

Identifiers

DOI
10.1088/0031-9155/54/8/002;
PII
S0031-9155(09)83421-0;

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
54
Journal Issue
8
Journal Page Range
p. 2277-2291
ISSN
0031-9155
CODEN
PHMBA7