Published 2010 | Version v1
Miscellaneous

Managing the backscatter component from the robotic arm of an a-Si EPID

  • 1. Central Coast Radiation Oncology Centre, New South Wales, Australia (Australia)
  • 2. Calvary Mater Hospital Newcastle, and the University of Newcastle, New South Wales, Australia (Australia)

Description

Full text: Backscatter from the robotic arm mechanism of an a-Si EPID in IMRT images was examined. Images corrected with a conventional flood field (FF) containing a backscatter component (BSC) from the robotic ann were compared with a BSC-free FF. A Yarian 21 EX linac (6 MV, 18 MV) was used. All images were acquired with two aS500 EPIDs, one R-arm and one E-arm. The BSC of an EPID image is the ratio of an image acquired with the EPID attached to the arm then detaching the arm from the EPID and acquiring the same image. A range of square field sizes from 2.5 x 2.5 cm to 27.5 x 27.5 cm were acquired and the BSC analyzed. The BSC of the FFs were also measured. A series of IMRT fields were acquired. Each field was corrected with a conventional FF and compared with a BSC-free FF. Figure I shows the magnitude of the BSC from each arm in the inplane for a 6x beam. Square fields above 16 x l6 cm (R-arm) and lO x 10 cm (E-arm) benefited from a conventional FF as it tended to cancel out the BSC in the acquired square field. The opposite was observed for smaller field sizes. A gamma analysis of the IMRT fields showed a FF correction containing a BSC reduces the effect of the arm in the final image. IMRT EPID images using conventional FFs have been shown to be less affected by backscatter from the robotic arm compared to BSC-free flood fields. (author)

Part of:
Program EPSM-ABEC 2010 Conference

Additional details

Publishing Information

Publisher
Australasian College of Physical Scientists and Engineers in Medicine
Imprint Place
Melbourne (Australia)
Imprint Title
Program EPSM-ABEC 2010 Conference
Imprint Pagination
62 p.
Journal Page Range
p. 146

Conference

Title
Conference abstract
Dates
5-9 Dec 2010
Place
Melbourne (Australia)

INIS

Country of Publication
Australia
Country of Input or Organization
Australia
INIS RN
43108923
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference, Non-conventional Literature
Descriptors DEI
BACKSCATTERING; DIAGNOSTIC TECHNIQUES; GAMMA SOURCES; IMAGE PROCESSING; IMAGES; NMR IMAGING; NUCLEAR MAGNETIC RESONANCE; RADIATION SOURCES; SCATTERING
Descriptors DEC
DIAGNOSTIC TECHNIQUES; MAGNETIC RESONANCE; PROCESSING; RADIATION SOURCES; RESONANCE; SCATTERING

Optional Information