Radiotherapy dose distribution in patients with metal implants - application of the Monte Carlo method

  • 1. Klinika radioterapie a onkologie, Fakultni nemocnice Kralovske Vinohrady, Prague (Czech Republic)
  • 2. Oddeleni klinicke hematologie, Fakultni nemocnice Kralovske Vinohrady, Prague (Czech Republic)

Description

The Monte Carlo method was applied to the calculation of the absorbed dose of a high-energy photon beam during radiotherapy of a tissue with a metallic material. Two materials, viz. iron and titanium, were used, and two radiation energies, 6 MV and 16 MV, were applied. The dose increase at the water-metal proximal junction did not exceed 30% and dropped rapidly with increasing distance from the junction. Except for the high-energy beam - iron combination, a shadow was apparent behind the metal. The higher energy level was associated with a lower dose inhomogeneity. This dose inhomogeneity, which is relatively low and limited in volume, can be additionally reduced in several ways, owing to which the indicated radiotherapy can be applied in such cases with a minimal risk

Additional details

Additional titles

Original title (Czech)
Distribuce davky radioterapie u pacientu s kovovymi implantaty - pouziti metody Monte Carlo

Publishing Information

Journal Title
Ceska Radiologie
Journal Volume
56
Journal Issue
1
Journal Page Range
p. 43-47
ISSN
1210-7883

Optional Information

Notes
English translation can be ordered from the Nuclear Information Center Zbraslav, 156 16 Prague 16 - Zbraslav, Czech Republic (e-mail: uji@uji.cz), at USD 10.- per standard page (1800 characters)