The Monte Carlo method was applied to the calculation of the absorbed dose of a high-energy photon beam during radiotherapy of a tissue with a metallic material. Two materials, viz. iron and titanium, were used, and two radiation energies, 6 MV and 16 MV, were applied. The dose increase at the water-metal proximal junction did not exceed 30% and dropped rapidly with increasing distance from the junction. Except for the high-energy beam - iron combination, a shadow was apparent behind the metal. The higher energy level was associated with a lower dose inhomogeneity. This dose inhomogeneity, which is relatively low and limited in volume, can be additionally reduced in several ways, owing to which the indicated radiotherapy can be applied in such cases with a minimal risk
Additional details
Additional titles
- Original title (Czech)
- Distribuce davky radioterapie u pacientu s kovovymi implantaty - pouziti metody Monte Carlo
Publishing Information
- Journal Title
- Ceska Radiologie
- Journal Volume
- 56
- Journal Issue
- 1
- Journal Page Range
- p. 43-47
- ISSN
- 1210-7883
INIS
- Country of Publication
- Czech Republic
- Country of Input or Organization
- Czech Republic
- INIS RN
- 33018594
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ANIMAL TISSUES; IMPLANTS; INTERFACES; IRON; MONTE CARLO METHOD; PHOTONS; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; RADIOTHERAPY; TITANIUM
- Descriptors DEC
- BODY; BOSONS; CALCULATION METHODS; DOSES; ELEMENTARY PARTICLES; ELEMENTS; MASSLESS PARTICLES; MEDICINE; METALS; NUCLEAR MEDICINE; RADIOLOGY; THERAPY; TRANSITION ELEMENTS
Optional Information
- Notes
- English translation can be ordered from the Nuclear Information Center Zbraslav, 156 16 Prague 16 - Zbraslav, Czech Republic (e-mail: uji@uji.cz), at USD 10.- per standard page (1800 characters)