Method of energy calibration of the TANSY neutron detectors

Description

A method to calibrate an array of scintillation neutron detectors, using a γ source, is presented. The count rate is measured as a function of high voltage at a given discrimination level. The obtained distribution is differentiated and a maximum value is determined which corresponds to the voltage at which the gamma peak passes through the discrimination level. By repeating the measurement at different discrimination levels the experimental dependence between the discrimination level and the high voltage is found as a straight line in a log-log diagram. Two calibration parameter for each detector are determined from a fit of these straight lines. A recalculation from the energy of the used γ source to any other energy is then possible and the obtained relation can be used to calculate discrimination levels and high voltages for each detector. Verification procedures are described. (authors)