Published June 1, 2011 | Version v1
Journal article

Determination of space charge distributions in highly segmented large volume HPGe detectors from capacitance-voltage measurements

  • 1. Institut fuer Kernphysik, Universitaet zu Koeln, Zuelpicher Strasse 77, 50937 Koeln (Germany)

Description

The space charge distribution of a large volume highly segmented HPGe detector was determined by a non-destructive capacitance-voltage measurement. The capacitances between the 36 segments and the core were measured simultaneously with a precision pulser which was implemented in the core preamplifier. The pulser measurement was compared to and validated by direct capacitance measurements. The three-dimensional doping profile was reconstructed using analytical and numerical methods. Consistent values for the impurity concentration in the range of 0.5 and 1.5x1010 cm-3 were obtained.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2011.02.109

Additional details

Identifiers

DOI
10.1016/j.nima.2011.02.109;
PII
S0168-9002(11)00591-2;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
640
Journal Issue
1
Journal Page Range
p. 176-184
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 2011 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.