Published September 1, 2008 | Version v1
Journal article

Synthetic single crystal diamond diodes for radiotherapy dosimetry

  • 1. INFN-Dipartimento di Ingegneria Meccanica, Universita di Roma 'Tor Vergata', Via del Politecnico 1, I-00133 Roma (Italy)
  • 2. U.O. Fisica Sanitaria, Ospedale San Filippo Neri, Via G. Martinotti 20, 00135 Roma (Italy)
  • 3. INFN-Dipartimento di Ingegneria Elettronica, Universita Roma 3, Via della Vasca Navale 84, 00146 Roma (Italy)
  • 4. Dipartimento di Ingegneria Elettronica, Universita Roma 3, Via della Vasca Navale 84, 00146 Roma (Italy)

Description

Synthetic single crystal diamonds in a p-type/intrinsic/metal structure were tested as dosimeters for radiotherapy. The devices have been analyzed by using 6 and 10 MV Bremsstrahlung X-ray beams and electron beams from 6 MeV up to 18 MeV from a CLINAC DHX Varian accelerator. All measurements have been performed in a water phantom and ionization chambers were used for calibration and comparison. The dosimeters were operated in photovoltaic regime with no external bias voltage applied. A few Gy pre-irradiation was performed in order to stabilize the device output, resulting in fluctuations sensitivity below ±0.5%. No dose rate dependence of the detector response was observed. Very good reproducibility and linearity were obtained as well

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2008.06.028

Additional details

Identifiers

DOI
10.1016/j.nima.2008.06.028;
PII
S0168-9002(08)00872-3;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
594
Journal Issue
2
Journal Page Range
p. 273-277
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 2008 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.