Published July 1977 | Version v1
Journal article

Assessment of a method for measuring serum thyroxine by radioimmunoassay, with use of polyethylene glycol precipitation

  • 1. General Hospital, St. John's, Newfoundland

Description

We assessed the efficacy of a new thyroxine radioimmunoassay kit (Abbott) in which polyethylene glycol is used to separate bound from free hormone. Mean serum thyroxine was 88 +- 15 (+-SD) μg/liter for 96 normal persons. Results for hypothyroid and hyperthyroid persons were clearly separated from those for normal individuals. Women taking oral contraceptive preparations showed variable increases in their serum thyroxine values. The coefficient of variation ranged from 1 to 3% within assay and from 5.4 to 11% among different assays. Excellent parallelism was demonstrated between thyroxine values estimated by this method and those obtained either by competitive protein binding or by a separate radioimmunoassay for the hormone

Additional details

Additional titles

Augmented title (English)
"1"2"5I tracer technique

Identifiers

Publishing Information

Journal Title
Clinical Chemistry
Journal Volume
23
Journal Issue
7
Series
Clin. Chem.
Journal Page Range
1333-1335