Characterization of OSL commercial dosimeters using a hand phantom, in standard beta radiation beams

  • 1. Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP) Sao Paulo, SP (Brazil)
  • 2. Fundacentro/MTE/Coordenacao de Higiene Ocupational, Sao Paulo, SP (Brazil)

Description

In this study, a hand phantom was developed and tested with Al2O3:C detectors (nanodots, Landauer) and the optically stimulated luminescence (OSL) technique. This paper shows the results of a characterization study of Al2O3:C detectors at the hand phantom, to simulate operations at a nuclear medicine service. The OSL detectors were exposed to standard beta radiation beans of the beta secondary standard system of the Calibration Laboratory at IPEN, with 90Sr + 90Y, 85Kr and 147Pm sources. The detectors were optically treated prior each reutilization. The results obtained show that for monitoring of workers exposed to beta radiation, the technique and the material are useful, but the energy dependence of the OSL response of Al2O3: C has to be taken into consideration. (author)

Files

41113438.pdf

Files (451.1 kB)

Name Size Download all
md5:b78f7b871bcc36442f31259a8223a6fb
451.1 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
[6 p.]
Report number
INIS-BR--7323

Conference

Title
International nuclear atlantic conference. Innovations in nuclear technology for a sustainable future; 16. Brazilian national meeting on reactor physics and thermal hydraulics; 9. Brazilian national meeting on nuclear applications; 1. Meeting on nuclear industry
Acronym
INAC 2009
Dates
27 Sep - 2 Oct 2009
Place
Rio de Janeiro, RJ (Brazil)

Optional Information

Notes
Published only in CD-Rom. Code: e02_1060_fullpaper.pdf