Description

The LiF:Mg,Cu,P phosphor is the most investigated phosphor in radiation dosimetry. Results on thermoluminescence of the epoxy based LiF:Mg,Cu,P dosimeters irradiated with gamma radiations are presented and compared with results of LiF:Mg,Cu,P powder. The glow curve structure of both LiF powder and dosimeter are same and only difference is found in the glow curve peak temperature. The LiF dosimeters were made from the 5012A and 5012B epoxy. The dosimeters had a mass of about 18 mg, 5.0 mm diameter and 0.5 mm thickness. The sensitivity variation of the dosimeters for exposure to 60Co gamma rays at different angles of incidence of the radiation is found to be within 4%. Its minimum detectable dose is about 3020 µGy. The epoxy based dosimeters withstand different environment and it can be used with general TL reader without need of any special design due to its small size and plane surface.

Additional details

Identifiers

Publishing Information

Journal Title
AIP Conference Proceedings
Journal Volume
1728
Journal Issue
1
Journal Page Range
vp.
ISSN
0094-243X
CODEN
APCPCS

Conference

Title
International conference on condensed matter and applied physics
Acronym
ICC 2015
Dates
30-31 Oct 2015
Place
Bikaner (India)

Optional Information

Notes
(c) 2016 Author(s)