Exposure dose in CT with special reference to exposure in CT-H 250

  • 1. Fujita-Gakuen Univ., Toyoake, Aichi (Japan). School of Medicine

Description

Exposure in a computed tomography, Hitachi CT-H 250 type, was discussed. Circumferential dose was measured by using elliptoidal phantoms (19 x 12, 12 x 11 cm) of Mix Dp. The maximum exposure dose by the large phantom was 2.9 rad and minimum dose was 0.23 rad. The maximum dose measured by the small phantom which was used by supposing the head of infants was 2.2 rad and the minimum dose was the same as that by the large phantom. Exposure dose out of flux (the neck, the abdomen, etc.), volume dose, and distribution of scattered x-ray in a room were also discussed. When an adaptor which restrained infant head was manufactured for trial, decrease of exposure dose was attempted by photographing with low voltage in a tube. Exposure dose was measured by the small phantom with low voltage of 120 kv, 100 kv, and 96 kv. As a result, the maximum dose of exposure decreased from 1740 to 1085 and 850 R which was under 50% of the first measurement value. There was not a significant difference in CT values and photographed pictures. (Serizawa, K.)

Additional details

Publishing Information

Journal Title
Rinsho Hoshasen
Journal Volume
21
Journal Issue
12
Series
Rinsho Hoshasen.
Journal Page Range
1073-1076

INIS

Country of Publication
Japan
Country of Input or Organization
Japan
INIS RN
8328238
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Descriptors DEI
RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; RADIATION PROTECTION; TOMOGRAPHY; X RADIATION
Descriptors DEC
ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; RADIATIONS