Heating induced near deep brain stimulation lead electrodes during magnetic resonance imaging with a 3 T transceive volume head coil
Creators
- 1. Center for Magnetic Resonance Research, University of Minnesota, 2021, 6th St SE, Minneapolis, MN 55455 (United States)
- 2. Department of Neurosurgery, University of Minnesota, D-429 Mayo, 420 Delaware Street SE, Minneapolis, MN 55455 (United States)
- 3. Division of Biostatistics, University of Minnesota A460 Mayo, MMC 303, 420 Delaware St SE, Minneapolis, MN 55455 (United States)
- 4. MR Safe Devices LLC, 14569 Grand Ave, Suite 168, Burnsville, MN 55306 (United States)
Description
Heating induced near deep brain stimulation (DBS) lead electrodes during magnetic resonance imaging with a 3 T transceive head coil was measured, modeled, and imaged in three cadaveric porcine heads (mean body weight = 85.47 ± 3.19 kg, mean head weight = 5.78 ± 0.32 kg). The effect of the placement of the extra-cranial portion of the DBS lead on the heating was investigated by looping the extra-cranial lead on the top, side, and back of the head, and placing it parallel to the coil's longitudinal axial direction. The heating was induced using a 641 s long turbo spin echo sequence with the mean whole head average specific absorption rate of 3.16 W kg−1. Temperatures were measured using fluoroptic probes at the scalp, first and second electrodes from the distal lead tip, and 6 mm distal from electrode 1 (T6mm). The heating was modeled using the maximum T6mm and imaged using a proton resonance frequency shift-based MR thermometry method. Results showed that the heating was significantly reduced when the extra-cranial lead was placed in the longitudinal direction compared to the other placements (peak temperature change = 1.5–3.2 °C versus 5.1–24.7 °C). Thermal modeling and MR thermometry may be used together to determine the heating and improve patient safety online. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/0031-9155/57/17/5651Additional details
Identifiers
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 57
- Journal Issue
- 17
- Journal Page Range
- p. 5651-5665
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 44049921
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BRAIN; ELECTRODES; HEATING; NMR IMAGING; PATIENTS; SIMULATION; SPIN ECHO; STIMULATION
- Descriptors DEC
- BODY; CENTRAL NERVOUS SYSTEM; DIAGNOSTIC TECHNIQUES; NERVOUS SYSTEM; ORGANS