Published August 1993 | Version v1
Journal article

Radiation dosimetry using junction field-effect transistor detectors

Creators

  • 1. Christian Medical Coll., Vellore (India)

Description

The use of junction field effect transistors (JFET) has been studied by connecting them in a bridge circuit. With a suitable back-up circuit, it was possible to measure doses as well as dose-rates. It was possible to alter the sensitivity of the JFET bridge by varying the biasing components of the JFET. Easy temperature compensation was also possible. However, response of the JFET to radiation showed energy dependency similar to that of semiconductor diodes. (author)

Additional details

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
38
Journal Issue
8
Journal Page Range
p. 1165-1172.
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
United Kingdom
INIS RN
24074670
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Descriptors DEI
COMPARATIVE EVALUATIONS; DOSIMETRY; ENERGY DEPENDENCE; JUNCTION DETECTORS; PERFORMANCE TESTING; RADIATION DOSES; RADIOBIOLOGY; SENSITIVITY
Descriptors DEC
BIOLOGY; EVALUATION; MEASURING INSTRUMENTS; RADIATION DETECTORS; SEMICONDUCTOR DETECTORS; TESTING