The Feasibility of Conformal Thermal Therapy with Transurethral Ultrasound Heating Applicators and MR Temperature Feedback
- 1. Sunnybrook and Women's College Health Sciences Centre, Toronto, Ontario (Canada)
- 2. Department of Medical Biophysics, University of Toronto, Toronto, Ontario (Canada)
Description
Transurethral thermal therapy offers a minimally invasive alternative for the treatment of prostate diseases including benign prostate hyperplasia (BPH) and prostate cancer. Accurate heating of a targeted region of the gland can be achieved through the use of a rotating directional heating source incorporating planar ultrasound transducers, and the implementation of active temperature feedback along the beam direction during heating provided by magnetic resonance (MR) thermometry. The performance of this control method with practical spatial, temporal, and temperature resolution (such as angular alignment, spatial resolution, update rate for temperature feedback (imaging time), and the presence of noise) for thermal feedback using a clinical 1.5 T MR scanner was investigated in simulations. As expected, the control algorithm was most sensitive to the presence of noise, with noticeable degradation in its performance above ±2 deg. C of temperature uncertainty. With respect to temporal resolution, acceptable performance was achieved at update rates of 5s or faster. The control algorithm was relatively insensitive to reduced spatial resolution due to the broad nature of the heating pattern produced by the heating applicator, this provides an opportunity to improve signal-to-noise ratio (SNR). The overall simulation results confirm that existing clinical 1.5T MR imagers are capable of providing adequate temperature feedback for transurethral thermal therapy without special pulse sequences or enhanced imaging hardware
Additional details
Identifiers
- DOI
- 10.1063/1.2205455;
Publishing Information
- Journal Title
- AIP Conference Proceedings
- Journal Volume
- 829
- Journal Issue
- 1
- Journal Page Range
- p. 146-150
- ISSN
- 0094-243X
- CODEN
- APCPCS
Conference
- Title
- 5. international symposium on therapeutic ultrasound
- Dates
- 27-29 Oct 2005
- Place
- Boston, MA (United States)
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 37103019
- Subject category
- S75: CONDENSED MATTER PHYSICS, SUPERCONDUCTIVITY AND SUPERFLUIDITY; S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ALGORITHMS; COMPUTERIZED SIMULATION; FEEDBACK; HEATING; NEOPLASMS; NMR IMAGING; NOISE; NUCLEAR MAGNETIC RESONANCE; PERFORMANCE; PROSTATE; PULSES; RADIOTHERAPY; SIGNAL-TO-NOISE RATIO; SPATIAL RESOLUTION; TEMPERATURE CONTROL; TRANSDUCERS; ULTRASONOGRAPHY
- Descriptors DEC
- BODY; CONTROL; DIAGNOSTIC TECHNIQUES; DIMENSIONLESS NUMBERS; DISEASES; GLANDS; MAGNETIC RESONANCE; MALE GENITALS; MATHEMATICAL LOGIC; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; RESOLUTION; RESONANCE; SIMULATION; THERAPY
Optional Information
- Notes
- (c) 2006 American Institute of Physics