Published October 2015 | Version v1
Miscellaneous

Monte Carlo characterization of the Gamma-Med Hdr plus Ir-192 brachytherapy source

  • 1. Universidad de Guanajuato, Division de Ciencias e Ingenierias, Av. Insurgentes 2354, 37150 Leon, Guanajuato (Mexico)
  • 2. IMSS, Unidad Medica de Alta Especialidad No. 1, Av. Adolfo Lopez Mateos 1813, 37340 Leon, Guanajuato (Mexico)

Description

Full text: The MCNP4C Monte Carlo code was used to simulate the dosimetry around the Gamma-Med Hdr Plus iridium-192 brachytherapy source in both air/vacuum and water environments. Dosimetry data in water was calculated and are presented into an away-along table. All dosimetric quantities recommended by the AAPM Task Group 43 report have been also calculated. These quantities are air kerma strength, dose rate constant, radial dose function and anisotropy function. The obtained data are compared to this source reference data, finding results in good agreement with them. In this study, recommendations of the AAPM TG-43U1 report have been followed and comply with the most recent AAPM and ESTRO physics committee recommendations about Monte Carlo techniques. The data in the present study complement published data and can be used as input in the Tps or as benchmark data to verify the results of the treatment planning systems as well as a means of comparison with other datasets from this source. (Author)

Availability note (English)

Available in abstract form only, full text entered in this record

Additional details

Publishing Information

Publisher
Sociedad Mexicana de Irradiacion y Dosimetria
Imprint Place
Mexico D. F. (Mexico)
Imprint Pagination
1 p.

Conference

Title
15. International Symposium on Solid State Dosimetry
Original Conference Title
15. Conferencia Internacional sobre Dosimetria de Estado Solido
Dates
26-30 Sep 2015
Place
Leon, Guanajuato (Mexico)