Monte Carlo characterization of the Gamma-Med Hdr plus Ir-192 brachytherapy source
Creators
- 1. Universidad de Guanajuato, Division de Ciencias e Ingenierias, Av. Insurgentes 2354, 37150 Leon, Guanajuato (Mexico)
- 2. IMSS, Unidad Medica de Alta Especialidad No. 1, Av. Adolfo Lopez Mateos 1813, 37340 Leon, Guanajuato (Mexico)
Description
Full text: The MCNP4C Monte Carlo code was used to simulate the dosimetry around the Gamma-Med Hdr Plus iridium-192 brachytherapy source in both air/vacuum and water environments. Dosimetry data in water was calculated and are presented into an away-along table. All dosimetric quantities recommended by the AAPM Task Group 43 report have been also calculated. These quantities are air kerma strength, dose rate constant, radial dose function and anisotropy function. The obtained data are compared to this source reference data, finding results in good agreement with them. In this study, recommendations of the AAPM TG-43U1 report have been followed and comply with the most recent AAPM and ESTRO physics committee recommendations about Monte Carlo techniques. The data in the present study complement published data and can be used as input in the Tps or as benchmark data to verify the results of the treatment planning systems as well as a means of comparison with other datasets from this source. (Author)
Availability note (English)
Available in abstract form only, full text entered in this recordAdditional details
Publishing Information
- Publisher
- Sociedad Mexicana de Irradiacion y Dosimetria
- Imprint Place
- Mexico D. F. (Mexico)
- Imprint Pagination
- 1 p.
Conference
- Title
- 15. International Symposium on Solid State Dosimetry
- Original Conference Title
- 15. Conferencia Internacional sobre Dosimetria de Estado Solido
- Dates
- 26-30 Sep 2015
- Place
- Leon, Guanajuato (Mexico)
INIS
- Country of Publication
- Mexico
- Country of Input or Organization
- Mexico
- INIS RN
- 47051812
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES; S97: MATHEMATICAL METHODS AND COMPUTING;
- Resource subtype / Literary indicator
- Conference, Non-conventional Literature
- Descriptors DEI
- AIR; ANISOTROPY; BENCHMARKS; BRACHYTHERAPY; COMPARATIVE EVALUATIONS; DATA; DATASETS; DOSE RATES; DOSES; DOSIMETRY; IRIDIUM 192; KERMA; MONTE CARLO METHOD; PLANNING; RECOMMENDATIONS; WATER
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; CALCULATION METHODS; DAYS LIVING RADIOISOTOPES; DOCUMENT TYPES; ELECTRON CAPTURE RADIOISOTOPES; EVALUATION; FLUIDS; GASES; HEAVY NUCLEI; HYDROGEN COMPOUNDS; INFORMATION; INTERNAL CONVERSION RADIOISOTOPES; IRIDIUM ISOTOPES; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; MEDICINE; MINUTES LIVING RADIOISOTOPES; NUCLEAR MEDICINE; NUCLEI; ODD-ODD NUCLEI; OXYGEN COMPOUNDS; RADIOISOTOPES; RADIOLOGY; RADIOTHERAPY; THERAPY; YEARS LIVING RADIOISOTOPES
Optional Information
- Funding organization
- National Council for Science and Technology (Mexico); International Centre for Theoretical Physics (Italy); Asesores en Proteccion Radiologica y Nuclear, S. C. (Mexico); DOSimetrics (Germany); RadMedical (Mexico); Convention and Visitors Bureau of Leon (Mexico); Tecnofisica, S. A. de C. V. (Mexico)