Published September 1990
| Version v1
Journal article
Position sensitive x-ray detector of arc shape with stripped anode
Description
Design and specifications of position-sensitive x-ray arc shape detector are presented. Quantum emission efficiency for CuKα is 25%. Detector with 150 mm radius of curvature in the process of its filling with argon (60%) isobutane (40%) mixture has angular aperture of 120 deg, providing angular resolution of 0.14 deg. Detector may be used for x-ray diffraction studied
Additional details
Additional titles
- Original title (Russian)
- Позиционно-чувствительный рентгеновский детектор дуговой формы с полосковым анодом
Publishing Information
- Journal Title
- Pribory i Tekhnika Ehksperimenta
- Journal Issue
- no.5
- Series
- Prib. Tekh. Ehksp.
- ISSN
- 0032-8162
- CODEN
- PRTEA
INIS
- Country of Publication
- Russian Federation
- Country of Input or Organization
- USSR
- INIS RN
- 23037067
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ARGON; BINARY MIXTURES; EFFICIENCY; POSITION SENSITIVE DETECTORS; SPATIAL RESOLUTION; X-RAY DETECTION; 2-METHYLPROPANE