Published May 2014 | Version v1
Journal article

Direct measurement of 60Co and 125I activity by the sum-peak method in PTKMR-BATAN

Description

The sum-peak method was successfully applied to the determination of the activity of 60Co and 125I sources measured by HPGe and well-type NaI(Tl) detectors in PTKMR-BATAN. The result of the 60Co activity agrees with the activity value measured by using a calibrated ionization chamber within a range of about 0.35–0.5% and the activity result of 125I agrees with the activity value measured using the photon–photon coincidence method within a range of 0.05–0.26%. - Highlights: • Sum-peak method applied to activity determination of 60Co and 125I. • Measured by well-type NaI(Tl) and HP(Ge) detectors. • Results agree (<0.5% and <0.3%) with those from calibrated ionization chambers

Availability note (English)

Available from http://dx.doi.org/10.1016/j.apradiso.2013.11.110

Additional details

Identifiers

DOI
10.1016/j.apradiso.2013.11.110;
PII
S0969-8043(13)00559-9;

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
87
Journal Page Range
p. 200-202
ISSN
0969-8043
CODEN
ARISEF

Conference

Title
19. international conference on radionuclide metrology and its applications
Dates
17-21 Jun 2013
Place
Antwerp (Belgium)

Optional Information

Copyright
Copyright (c) 2014 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.