A novel APD-based detector module for multi-modality PET/SPECT/CT scanners
Description
The lack of anatomical information in SPECT and PET images is one of the major factors limiting the ability to localize and accurately quantify radionuclide uptake in small regions of interest. This problem could be resolved by using multi-modality scanners having the capability to acquire anatomical and functional images simultaneously. The feasibility of a novel detector suitable for measuring high-energy annihilation radiation in PET, medium-energy γ-rays in SPECT and low-energy X-rays in transmission CT is demonstrated and its performance is evaluated for potential use in multi-modality PET/SPECT/CT imaging. The proposed detector consists of a thin CsI(Tl) scintillator sitting on top of a deep GSO/LSO pair read out by an avalanche photodiode. The GSO/LOS pair provides depth-of-interaction information for 511 keV detection in PET, while the thin CsI(Tl) that is essentially transparent to annihilation radiation is used for detecting lower energy X- and γ-rays. The detector performance is compared to that of an LSO/YSO phoswich. Although the implementation of the proposed GSO/LSO/CsI(Tl) detector raises special problems that increase complexity, it generally outperforms the LSO/YSO phoswich for simultaneous PET, SPECT and CT imaging
Additional details
Publishing Information
- Journal Title
- IEEE Transactions on Nuclear Science
- Journal Volume
- 46
- Journal Issue
- 3Pt2
- Journal Page Range
- p. 479-484
- ISSN
- 0018-9499
- CODEN
- IETNAE
Conference
- Title
- 1998 IEEE nuclear science symposium and medical imaging conference
- Dates
- 10-12 Nov 1998
- Place
- Toronto (Canada)
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 31001185
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- EMISSION COMPUTED TOMOGRAPHY; GAMMA CAMERAS; IMAGE PROCESSING; OPTIMIZATION; PERFORMANCE TESTING; SCINTILLATION COUNTERS
- Descriptors DEC
- CAMERAS; COMPUTERIZED TOMOGRAPHY; MEASURING INSTRUMENTS; RADIATION DETECTORS; TESTING; TOMOGRAPHY