Published February 1986
| Version v1
Journal article
Noise distribution due to emission and transmission statistics in positron emission tomography
- 1. UBC/TRIUMF PET Program, The Univ. of British Columbia, 2211 Wesbrook Mall, Vancouver, British Columbia V6T 2B5
Description
Noise contributions to PET image data are due mainly to the limited counting statistics in the profiles of both emission and transmission scans. Statistical variance ''images'' were obtained with the UBC/TRIUMF PETT VI tomograph representing noise due to the corresponding emission and transmission scans. Phantom measurement variances were compared with those predicted analytically and found to be in close agreement. Variance contributions due to transmission noise are found to be most pronounced towards the centre of the image
Additional details
Publishing Information
- Journal Title
- IEEE Trans. Nucl. Sci.
- Journal Volume
- NS-33
- Journal Issue
- 1
- Series
- IEEE Trans. Nucl. Sci.
- Journal Page Range
- vp.
- ISSN
- 0018-9499
- CODEN
- IETNA
Conference
- Title
- IEEE nuclear science symposium.
- Dates
- 23-25 Oct 1985.
- Place
- San Francisco, CA (USA).
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 17068497
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- IMAGE PROCESSING; NOISE; PHANTOMS; POSITRON COMPUTED TOMOGRAPHY; STATISTICS; TRIUMF CYCLOTRON
- Descriptors DEC
- ACCELERATORS; COMPUTERIZED TOMOGRAPHY; CYCLIC ACCELERATORS; CYCLOTRONS; EMISSION COMPUTED TOMOGRAPHY; ISOCHRONOUS CYCLOTRONS; MATHEMATICS; MOCKUP; STRUCTURAL MODELS; TOMOGRAPHY
Optional Information
- Secondary number(s)
- CONF-851009--.