A deformable phantom was developed to aid in quality assurance for dynamic imaging and targeting techniques in radiation therapy. Made of simple materials combined with standard components for imaging and motion experiments, this phantom can be relatively easily constructed and used for both diagnostic imaging and dosimetry. Repeat imaging studies indicate that the phantom meets criteria of relative attenuation, deformation, and reproducibility of configuration, necessary for quality assurance of radiographic and tomographic targeting
Additional details
Identifiers
- DOI
- 10.1118/1.2400612;
Publishing Information
- Journal Title
- Medical Physics
- Journal Volume
- 34
- Journal Issue
- 1
- Journal Page Range
- p. 199-201
- ISSN
- 0094-2405
- CODEN
- MPHYA6
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 38026900
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- COMPUTERIZED TOMOGRAPHY; DEFORMATION; DOSIMETRY; PHANTOMS; QUALITY ASSURANCE; RADIOTHERAPY; SIMULATION
- Descriptors DEC
- DIAGNOSTIC TECHNIQUES; MEDICINE; MOCKUP; NUCLEAR MEDICINE; RADIOLOGY; STRUCTURAL MODELS; THERAPY; TOMOGRAPHY
Optional Information
- Notes
- (c) 2007 American Association of Physicists in Medicine