Dead-time analysis of digital spectrometers

Description

This paper studies the impact that the outcome of new digital architectures has brought to dead time of spectrometers. Special emphasis is given to the performance comparison between conventional spectrometers and the most important architectures usually adopted in the digital solutions. Dead-time analysis is accomplished through a dedicated numerical simulation application developed with the purpose of optimizing the digital architecture performance taking into account parameters such as the incident rate of events, the existence of pulse pile-up or the complexity of the processing algorithm used during the digital pulse processing. Results of simulations using a real time multiprocessor platform (Digital Pulse Processing-M8) are presented and design options are justified accordingly

Additional details

Identifiers

DOI
10.1016/j.nima.2003.11.389;
PII
S0168900203033151;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
522
Journal Issue
3
Journal Page Range
p. 487-494
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 2003 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.