Study on Oneself Developed to Apparatus Position of Measurement of BMD in the Distal Radius
- 1. Dept. of Radiology, Chungnam National University Hospital, Daejeon (Korea, Republic of)
- 2. Dept. of Radiology, Eulji University Hospital, Daejeon (Korea, Republic of)
- 3. Dept. of Radiology, Konyang University Hospital, Daejeon (Korea, Republic of)
- 4. Dept. of Radiological Technology, Daewon University College, Jecheon (Korea, Republic of)
Description
The aim of this study was to evaluate the difference of bone mineral density according to distal radius rotation and to develop the supporting tool to measure rotation angles. CT scanning and the measurement of BMD by DXA of the appropriate position of the forearm were performed on 20 males. Twenty healthy volunteers without any history of operations, anomalies, or trauma were enrolled. The CT scan was used to evaluate the cross sectional structure and the rotation angle on the horizontal plane of the distal radius. The rotational angle was measured by the m-view program on the PACS monitor. The DXA was used in 20 dried radii of cadaveric specimens in pronation and supination with five and ten degrees, respectively, including a neutral position (zero degrees) to evaluate the changes of BMD according to the rotation. The mean rotation angle of the distal radius on CT was 7.4 degrees of supination in 16 cases (80%), 3.3 degrees of pronation in three cases (15%), and zero degree of neutral in one case (9%), respectively. The total average rotation angle in 20 people was 5.4 degrees of supination. In the cadaveric study, the BMD of the distal radius was different according to the rotational angles. The lowest BMD was obtained at 3.3 degrees of supination. In the case of the measurement of BMD in the distal radius with a neutral position, the rotational angle of the distal radius is close to supination. Pronation is needed for the constant measurement of BMD in the distal radius with the rotation angle measuring at the lowest BMD and about five degrees of pronation of the distal radius is recommended.
Additional details
Publishing Information
- Journal Title
- Journal of the Korean Society of Radiological Technology
- Journal Volume
- 32
- Journal Issue
- 4
- Series
- 9 refs, 11 figs, 2 tabs
- Journal Page Range
- p. 419-426
- ISSN
- 1226-2854
INIS
- Country of Publication
- Korea, Republic of
- Country of Input or Organization
- Korea, Republic of
- INIS RN
- 44015395
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- COMPUTERIZED TOMOGRAPHY; CROSS SECTIONS; EQUIPMENT; MINERALS; SKELETON
- Descriptors DEC
- BODY; DIAGNOSTIC TECHNIQUES; ORGANS; TOMOGRAPHY