Published 1982 | Version v1
Journal article

Evaluation of a linear position sensitive neutron detector for powder spectroscopy

  • 1. Univ. of Rhode Island, Kingston

Description

A one meter long position sensitive detector was acquired from Harshaw Corp. for a small angle neutron camera operated by the University of Rhode Island at the 2 MW reactor of the Rhode Island Nuclear Science Center. In the course of testing the detector system, to determine linearity, resolution, stability, and efficiency, it was employed in a powder spectrometer. Once geometrical effects are properly taken care of, it appears that the present detector, though not optimized for powder diffraction, permits collection of data essentially equivalent to that taken with the original step scanned detector though 15 times faster

Additional details

Publishing Information

Journal Title
AIP Conf. Proc.
Journal Issue
no.89
Series
AIP Conf. Proc.
ISSN
0094-243X
CODEN
APCPC

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
17028621
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Numerical Data
Descriptors DEI
EFFICIENCY; EXPERIMENTAL DATA; NEUTRON SPECTROSCOPY; PERFORMANCE; POSITION SENSITIVE DETECTORS; RESOLUTION; STABILITY
Descriptors DEC
DATA; INFORMATION; MEASURING INSTRUMENTS; NUMERICAL DATA; RADIATION DETECTORS; SPECTROSCOPY