An investigation of the accuracy of EGG-trigged cardio computer tomography using a moving phantom

  • 1. Bonn Univ. (Germany, F.R.). Radiologische Klinik
  • 2. Bonn Univ. (Germany, F.R.). Inst. fuer Strahlenbiologie

Description

Triggered CT-cuts of a model of the left ventricle were obtained in order to evaluate the accuracy of the procedure for measuring the cross sectional area and volume. The model was a balloon placed in water and pulsating with a variable frequency. The volume was estimated by an approximation method which is used in clinical practice, making use of the greatest possible cross sectional area and its diameter. It was concluded that the ejection fraction could be calculated to an accuracy of +- 11%. (orig.)

Abstract (German)

An einem Modell des linken Ventrikels wurden getriggerte CT-Aufnahmen durchgefuehrt mit dem Ziel, die Genauigkeit abzuschaetzen, mit der bei diesem Abbildungsverfahren Querschnittsflaechen bzw. Volumina ermittelt werden koennen. Als Modell wurde ein mit variabler Frequenz pulsierender, in Wasser gelagerter Ballon verwendet. Das Volumen wurde aus einer in der klinischen Praxis angewandten Naeherungsformel ermittelt, in die eine - moeglichst grosse - Querschnittsflaeche sowie deren groesster Durchmesser eingehen. Es ergab sich bei der Ermittlung der Ejektionsfraktion eine Messgenauigkeit von +- 11%. (orig.)

Additional details

Additional titles

Original title (German)
Untersuchungen zur Messgenauigkeit bei der EKG-getriggerten Kardiocomputertomographie an einem bewegten Phantom

Publishing Information

Journal Title
Fortschr. Geb. Roentgenstr. Nuklearmed.
Journal Volume
137
Journal Issue
5
Series
CODEN: FGRNA.;Fortschr. Geb. Roentgenstr. Nuklearmed.
Journal Page Range
544-548
ISSN
0015-8151

Optional Information

Notes
With 5 figs.