The effects of zero-filling in magnetic resonance imaging
- 1. California Univ., Los Angeles (USA). Dept. of Radiological Sciences
- 2. California Univ., Los Angeles (USA). School of Medicine
- 3. Innovision Diagnostic Imaging Corporation, Bellport (USA)
Description
Fast Fourier transform algorithms used by most MR manufacturers for two-dimensional Fourier transform (2D-FT) magnetic resonance imaging require a power of two number of sample points along the phase and frequency encoding directions. Zero-filling is a technique that allows smaller sample sizes to be used by adding zeroes to bring the data set to a full power of two. The effect of variable amounts of zero-filling on scan time, image signal-to-noise, contrast, and spatial resolution was empirically studied using phantoms and normal volunteers. Zero-filling allows a dramatic reduction in the MR scan time at the expense of decreased spatial resolution and increased truncation errors along the phase encoding axis. Unlike most other techniques for shortening 2D-FT scan times, zero-filling largely preserves the signal-to-noise response. 9 refs.; 4 figs.; 1 table
Additional details
Publishing Information
- Journal Title
- Journal of Medical Imaging
- Journal Volume
- 2
- Journal Issue
- 3
- Series
- J. Med. Imaging.
- Journal Page Range
- 184-190
- ISSN
- 0920-5497
- CODEN
- JMEIE
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 20026361
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ALGORITHMS; FOURIER TRANSFORMATION; NMR IMAGING; SPATIAL RESOLUTION; TWO-DIMENSIONAL CALCULATIONS
- Descriptors DEC
- DIAGNOSTIC TECHNIQUES; INTEGRAL TRANSFORMATIONS; RESOLUTION; TRANSFORMATIONS