Published 2015 | Version v1
Miscellaneous Open

Development of conditions radiation in computed tomography - CT

  • 1. Instituto de Radioprotecao e Dosimetria (IRD/CNEN-RJ), Rio de Janeiro, RJ (Brazil). Lab. Nacional de Metrologia das Radiacoes Ionizantes

Description

The LNMRI is implementing the TC qualities, used in chambers calibration used in hospitals and clinics. Preliminary studies results showed a variation for RQT 9 Quality of 1,350 x 107 and 1,343 x 107 Gy/C within 1.61 and 1.70% expanded uncertainties evaluated. (author)

Files

47080163.pdf

Files (327.0 kB)

Name Size Download all
md5:dfa2be891fa03496deda5a355f2dfd4f
327.0 kB Preview Download

Additional details

Additional titles

Original title (Portuguese)
Desenvolvimento das condicoes de radiacao em tomografia computadorizada - TC

Publishing Information

Imprint Pagination
3 p.
Report number
INIS-BR--16425

Conference

Title
2. Brazilian congress on metrology of ionizing radiations
Acronym
CMRI 2015
Dates
29 Nov - 4 Dec 2015
Place
Bento Goncalves, RS (Brazil)