Published October 1995
| Version v1
Report
Open
Dosimetry system for heavy-ion radiotherapy
- 1. National Inst. of Radiological Sciences, Chiba (Japan)
Description
This paper describes a dosimetry system for the heavy-ion radiation therapy installed at HIMAC facility. Daily check of the dose monitor, a species of the accelerated particles, their energy are carried out every morning. And the dose calibration of the monitor for each patient condition are also carried out every morning. The variations of the monitor responses were less than 2% during 3 months of treatment term. In order to shorten the time for the check, a simplified dosimetric system for the dose calibration of the monitor was installed from this autumn. (author)
Files
27044971.pdf
Files
(77.1 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:6a0a5bc7d50730de730fdbe58cfd1f0b
|
77.1 kB | Preview Download |
Additional details
Publishing Information
- Imprint Title
- Proceedings of the 10th symposium on accelerator science and technology
- Imprint Pagination
- 490 p.
- Journal Page Range
- p. 448-450.
- Report number
- JAERI-Conf--95-021
Conference
- Title
- 10. symposium on accelerator science and technology.
- Dates
- 25-27 Oct 1995.
- Place
- Hitachinaka, Ibaraki (Japan).
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 27044971
- Subject category
- S43: PARTICLE ACCELERATORS;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BEAM MONITORING; CALIBRATION; CARBON IONS; DOSIMETRY; HEAVY IONS; HIMAC ACCELERATOR; PATIENTS; RADIATION DOSES; RADIOTHERAPY; THERAPEUTIC USES
- Descriptors DEC
- ACCELERATORS; ANIMALS; CHARGED PARTICLES; CYCLIC ACCELERATORS; HEAVY ION ACCELERATORS; IONS; MAMMALS; MAN; MEDICINE; MONITORING; PRIMATES; SYNCHROTRONS; THERAPY; USES; VERTEBRATES