Published 2007 | Version v1
Journal article

Radiation safety survey on a flattening filter-free medical accelerator

  • 1. The Univ. of Texas MD Anderson Cancer Center, Holcombe Blvd., Houston, TX 77030 (United States)

Description

Dose rates at several locations outside a treatment room were measured for 6 and 18 MV photon beams from a Varian Clinac 21EX accelerator operated with and without a flattening filter. Also, dose rates in the treatment room due to activation were measured at 18 MV. An analysis of the measured data is presented. The results suggest that substantial reduction in doses outside the treatment room and lower activation can be achieved with a flattening-filter free accelerator. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/ncm179

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
124
Journal Issue
2
Journal Page Range
p. 187-190
ISSN
0144-8420

INIS

Country of Publication
United Kingdom
Country of Input or Organization
France
INIS RN
44022164
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY; S43: PARTICLE ACCELERATORS;
Descriptors DEI
ACCELERATORS; DESIGN; DOSE RATES; PHOTON BEAMS; RADIATION DOSES; RADIATION MONITORING; RADIATION PROTECTION
Descriptors DEC
BEAMS; DOSES; MONITORING

Optional Information

Notes
14 refs