Radiation safety survey on a flattening filter-free medical accelerator
- 1. The Univ. of Texas MD Anderson Cancer Center, Holcombe Blvd., Houston, TX 77030 (United States)
Description
Dose rates at several locations outside a treatment room were measured for 6 and 18 MV photon beams from a Varian Clinac 21EX accelerator operated with and without a flattening filter. Also, dose rates in the treatment room due to activation were measured at 18 MV. An analysis of the measured data is presented. The results suggest that substantial reduction in doses outside the treatment room and lower activation can be achieved with a flattening-filter free accelerator. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1093/rpd/ncm179Additional details
Identifiers
- DOI
- 10.1093/rpd/ncm179;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 124
- Journal Issue
- 2
- Journal Page Range
- p. 187-190
- ISSN
- 0144-8420
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 44022164
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S43: PARTICLE ACCELERATORS;
- Descriptors DEI
- ACCELERATORS; DESIGN; DOSE RATES; PHOTON BEAMS; RADIATION DOSES; RADIATION MONITORING; RADIATION PROTECTION
- Descriptors DEC
- BEAMS; DOSES; MONITORING
Optional Information
- Notes
- 14 refs