Verification and optimization of the optical parameters for slow extracted beam
Creators
- 1. National Institutes for Quantum and Radiological Science and Technology, Chiba (Japan)
- 2. Kanagawa Cancer Center Hospital, Yokohama, Kanagawa (Japan)
Description
Control of the beam size and position for three-dimensional pencil-beam scanning is important. We have developed a simple procedure for the optimization of the optical parameters and the beam alignment. Verification of the beam optics is performed by the measured beam size and variation of the beam position along the beam transfer line. Optimization of the beam optics is performed with the measured optical parameters at the extraction point. After the optimization, beamline tuning is performed with steering magnet and fluorescent screen. The misalignment between the beam position and reference axis in the treatment room is checked by the verification system, which consists of the screen monitor and an acrylic phantom. If the beam position is deviated from the reference axis, two pairs of steering magnets that are placed downstream of the beam transport line will correct the beam. These adjustments are iterated until the deviations for 11 energies of the beam are within 0.5 mm of the reference axis. To indicate the success of our procedure, we performed experiment at the Kanagawa Cancer Center using our procedure. (author)
Additional details
Identifiers
Publishing Information
- Imprint Title
- Proceedings of the 14th annual meeting of Particle Accelerator Society of Japan
- Imprint Pagination
- [1427 p.]
- Journal Page Range
- p. 1254-1256
Conference
- Title
- 14. annual meeting of Particle Accelerator Society of Japan
- Acronym
- PASJ2017
- Dates
- 1-3 Aug 2017
- Place
- Sapporo, Hokkaido (Japan)
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 49048516
- Subject category
- S43: PARTICLE ACCELERATORS;
- Resource subtype / Literary indicator
- Conference, Non-conventional Literature
- Descriptors DEI
- ALIGNMENT; BEAM EMITTANCE; BEAM EXTRACTION; BEAM MONITORS; BEAM OPTICS; BEAM POSITION; BEAM TRANSPORT; CHARGE-COUPLED DEVICES; ELECTROMAGNETS; EXTERNAL BEAM RADIATION THERAPY; MAGNETIC FIELDS; PARTICLE BEAMS; SYNCHROTRONS; TRANSFER MATRIX METHOD
- Descriptors DEC
- ACCELERATORS; BEAMS; CALCULATION METHODS; CYCLIC ACCELERATORS; ELECTRICAL EQUIPMENT; EQUIPMENT; MAGNETS; MEASURING INSTRUMENTS; MEDICINE; MONITORS; NUCLEAR MEDICINE; RADIOLOGY; RADIOTHERAPY; SEMICONDUCTOR DEVICES; THERAPY
Optional Information
- Notes
- 7 refs., 6 figs.