Published June 2003 | Version v1
Journal article

Use of BANG registered polymer gel for dose measurements in a 68 MeV proton beam

  • 1. Hahn-Meitner Institut, Augentumortherapie, Glienicker Strasse 100, D-14109 Berlin (Germany)
  • 2. Technische Fachhochschule Berlin, Medizinphysik, Luxemburger Strasse 10, D-13353 Berlin (Germany)
  • 3. Humboldt Universitaet zu Berlin, Charite Campus Virchow-Klinikum, Klinik fuer Strahlenheilkunde, Augustenburger Platz 1, D-13353 Berlin (Germany)
  • 4. Humboldt Universitaet zu Berlin, Charite Campus Mitte, Klinik fuer Strahlentherapie, Schumannstrasse 20/21, D-10117 Berlin (Germany)

Description

BANG polymer gel dosimetry using magnetic resonance imaging (MRI) was applied to an ophthalmologic 68 MeV proton beam. The object was to examine the use of BANG gel for the verification of proton fields in eye tumor therapy and to explore the applicability of polymer gel dosimetry in proton therapy under practical aspects. The gel phantoms were irradiated with monoenergetic and modulated proton beams. MRI analysis was carried out at clinical 1.5 and 3 T MR scanners. At constant LET, results show a linear relationship between spin-spin relaxation rates and dose. However, depth dose curves in BANG gel reveal a quenching of the Bragg maximum due to LET effects. The dose response of the gel for monoenergetic protons and spread-out depth dose distributions can be calculated based on ionization chamber measurements. Experiment and calculations show good agreement and indicate that BANG polymer gels might become a valuable tool in proton therapy quality assurance

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
30
Journal Issue
6
Journal Page Range
p. 1235-1240
ISSN
0094-2405
CODEN
MPHYA6

Optional Information

Notes