Published August 1992 | Version v1
Journal article

Quality control of SPECT

  • 1. PUMC Hospital, CAMS, Beijing (China)

Description

Point, line and sensitivity phantoms were designed. Jaszczak phantom was selected as the tomographic uniformity and total performance phantom. Six parameters were proposed to evaluate the performance of SPECT. The results obtained from the omega 500/Mcs 560 SPECT indicated that the electronic performance was acceptable but the mechanical performance was greatly degraded due to the mechanical wearing and aging of this equipment

Additional details

Publishing Information

Journal Title
Chinese Journal of Nuclear Medicine
Journal Volume
12
Journal Issue
3
Journal Page Range
p. 171-172.
ISSN
0253-9780
CODEN
CITCDE