Published August 2002 | Version v1
Journal article

Characterization and validation of the gamma-scintigraphic method for determining liquid holdup in foam

Description

Characterization and validation protocols for quantifying liquid content in a protein foam by gamma-scintigraphy were developed. Spatial resolution, count resolution and volume resolution values were measured to determine the analytical limitations of the apparatus. Scintigraphic measures of volume, in both liquid and dispersed media, correlated well with physical measures. Ionic interaction between [99mTc]pertechnetate and bovine serum albumin was undetectable by gel permeation. These results confirm that gamma-scintigraphy can be used to quantify liquid holdup in protein foams of pharmaceutical interest

Additional details

Identifiers

PII
S0969804302001057;

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
57
Journal Issue
2
Journal Page Range
p. 243-255
ISSN
0969-8043
CODEN
ARISEF

Optional Information

Copyright
Copyright (c) 2002 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.