Published 2006 | Version v1
Journal article

Sensitivity loss of LiF: Mg,Cu,P thermoluminescence dosemeters caused by oven annealing

  • 1. Univ. of Veterinary Medicine Hannover, Inst. for General Radiology and Medical Physics, Bischofsholer Damm 15, D-30173 Hannover (Germany)
  • 2. Univ. of Ulm, Dept. of Radiotherapy, Robert-Koch-Strasse 6, D-89081 Ulm (Germany)

Description

LiF:Mg,Cu,P is a very sensitive thermoluminescence material that can be used for personal dosimetry in radiology. But if this material is heated too much during annealing or reading it quickly loses its sensitivity. This study shows that an annealing oven in wide use can cause thermal damage to the dosemeters owing to inhomogeneous temperature distribution in the annealing tray. At annealing temperatures >240 deg. C, differences of only 1 deg. C lead to significant losses of sensitivity. Therefore, it is necessary to measure the distribution of temperature in the annealing tray for correct placement of the dosemeters in the tray. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/ncl008

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
121
Journal Issue
2
Journal Page Range
p. 195-201
ISSN
0144-8420

Optional Information

Notes
11 refs